Radical-7 Appendix: Best Practices for Comparisons to Reference Measurements
www.masimo.com 141 Masimo
Secure sensor cable to patient’s arm with tape.
Instruct subjects to remain still without any sensor movement.
Connect sensor cable to device without pulling on sensor.
Wait until SpHb measurement is stable (one minute) before recording SpHb
values.
Note any Low SIQ message that displays during measurement.
Blood sampling and laboratory analysis:
Record site of blood draw.
Document exact time blood withdrawn from patient.
When taking blood sample from arterial line, remove adequate dead space to
ensure blood sample is not diluted with solution to keep arterial line open.
When IV line set-up includes stopcocks, ensure that no fluid is infused
through the line from which you are taking the blood. Draw blood directly
into the vacutainer, if possible.
Mix blood carefully after filling the vacutainer by rotating it gently at least 10
times. (A rotation is one turn of the tube upside down and then right side
up.)
All blood samples must be analyzed on the same calibrated laboratory
hematology analyzer (Coulter, Sysmex, or equivalent) to avoid variation
induced by the use of multiple laboratory devices.