User's Manual
Table Of Contents
- About this Manual
- Product Description, Indications for Use, Contraindications, and Features
- Safety Information, Warnings, and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology (SET)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring (RAM) Technology
- Chapter 2: Radical-7 Descriptions
- Chapter 3: Setup
- Chapter 4: Operation
- Using the Touchscreen and Buttons
- Using Screen Lock
- Using the Home Button
- Navigating the Radical-7
- About the Display View
- Sensitivity Modes Overview
- Changing Sensitivity Modes
- Accessing the Main Menu
- Navigating the Main Menu
- Parameter Settings
- Sounds
- Device Settings
- Trends
- About
- Chapter 5: Profiles
- Chapter 6: Alarms and Messages
- About Alarms
- Silencing the Alarms
- Adaptive Threshold Alarm (ATA) Feature
- 3D Alarms
- Messages
- Replace Sensor Message
- Replace Cable Message
- Replace Adhesive Sensor Message
- Incompatible Sensor Message
- Incompatible Adhesive Sensor Message
- No Adhesive Sensor Connected Message
- Interference Detected Message
- SpO2 Only Mode Message
- Low Battery Message
- Low Perfusion Index Message
- Low Signal IQ Message
- Low SpCO SIQ Message
- Low SpMet SIQ Message
- Low SpHb SIQ Message
- Speaker Failure Message
- No Cable Connected Message
- No Sensor Connected Message
- Pulse Search Message
- Sensor Initializing Message
- Sensor Off Patient Message
- Incompatible Cable Message
- Chapter 7: Troubleshooting
- Chapter 8: Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Trending
- Alarms
- Display Indicators
- Compliance
- Output Interface
- Wireless Radio (If Installed)
- Serial Interface Specifications
- Serial Interface Setup
- Analog Output and Nurse Call Specifications
- Symbols
- Country Codes (FCC and EU)
- Citations
- Chapter 9: Service and Maintenance
- Appendix: Best Practices for Comparisons to Reference Measurements
- Index
Radical-7 Chapter 9: Service and Maintenance
www.masimo.com 135 Masimo
Nurse Call Test
To conduct a Nurse Call test
1. Disconnect any patient cables, sensors, or accessories from the Radical-7. Turn off
the Radical-7 and then turn on again.
2. Ensure that there are no audible alarms and that the Audible Alarm feature is not
set to silenced.
3. Verify the Nurse Call polarity is set to normal (default).
4. Connect the common lead of a digital multi-meter to the pin 12 (Nurse Call
Common) of the Analog Output connector on the Radical-7. Connect the positive
lead of the multi-meter to pin 6 (Nurse Call - Normally Open) of the Analog Output
connector and measure that the resistance is greater than 1 MW (open circuit).
5. Trigger an alarm on the monitor (for example, by disconnecting a sensor after it
was measuring data). Verify that the resistance is less than 35 ohms.
Analog Output Test
To conduct an Analog Output test
1. Disconnect any patient cables, sensors, or accessories from the Radical-7. Turn off
the Radical-7 and then turn on again.
2. Connect the common lead of a digital voltmeter to the pin 2 (Ground) of the
analog output connector on the Radical-7. Connect the positive lead of the
voltmeter to pin 9 (Analog 1) of the analog output connector.
3. On the device output screen, on the analog 1 option, select 0V Output. See Device
Output on page 80.
4. Verify that the voltmeter measures a voltage of approximately 0V.
5. Change the analog 1 option to 1V Output.
6. Verify that the voltmeter measures a voltage of approximately 1.0V.
7. Repeat steps 5 and 6, with the positive lead of the voltmeter connected to pin 15
(analog 2). See Serial Interface Specifications on page 122.
8. Connect a patient cable and sensor and verify that the voltage on pins 9 and 15
are between 0V and 1.0V while measuring a saturation and pulse rate.
Battery Test
To conduct a Battery test
1. Fully charge the Radical-7 by placing the Handheld into the Docking Station and
then connect the AC power.
2. Verify that the Handheld Battery Charging indicator is illuminated.
3. When the Radical-7 is fully charged, the Handheld Battery Charging indicator
turns off.
4. Turn on the Radical-7 on and verify that the Battery indicator shows a full charge.