User's Manual

Table Of Contents
www.masimo.com 131 Masimo
Chapter 9: Service and Maintenance
The following chapter contains information about cleaning, battery operation, performance
verification, service, repair, and warranty.
Cleaning
Under normal operation, no internal adjustment or recalibration is required. Safety tests and
internal adjustments should be done by qualified personnel only. Safety checks should be
performed at regular intervals or in accordance with local and governmental regulations.
Safety Information, Warnings, and Cautions on page 11.
The Radical-7 Pulse CO-Oximeter is a reusable instrument. The instrument is supplied and
used non-sterile.
To surface clean the Radical-7
The outer surface of the instrument can be cleaned with a soft cloth dampened
with a mild detergent and warm water solution.
Do not allow liquids to enter the interior of the instrument.
The outer surface of the instrument can also be wiped down using any of the
following solvents:
Cidex Plus (3.4% glutaraldehyde)
10% bleach solution
70% isopropyl alcohol solution
Using the recommended cleaning solutions on the touchscreen panel will not affect the
performance of the Handheld.
Battery Operation and Maintenance
The Radical-7 Handheld includes a lithium ion rechargeable battery. The Radical-7 Docking
Station may include the optional 6.5 amp-hour nickel metal hydride rechargeable battery.
Before using the Radical-7 as a Handheld or as a transport monitor, the Handheld
rechargeable battery and the optional Docking Station rechargeable battery must be fully
charged.
To charge the Handheld rechargeable battery and the Docking Station rechargeable
battery
1. Attach the Handheld to the Docking Station.
2. Connect the Docking Station to AC power.
3. Verify that the batteries are charging.
The Docking Station Battery Charging indicator momentarily flashes and
then remains illuminated while the batteries are actively charging.
If the internal battery temperature exceeds recommended operating
conditions for proper battery charging, the Handheld Battery Charging