Radical-7 Chapter 8: Specifications
www.masimo.com 119 Masimo
Alarms
Parameter Alarm Range
SpO2 1% to 99%
SpCO 1% to 98%
SpMet 0.1% to 99.5%
SpHb 1.0 g/dL to 24.5 g/dL
RR 5 breaths per minute to 69 breaths per
minute
PI 0.03% to 19%
PVI 1% to 99%
Pulse Rate 30 bpm to 235 bpm
SpOC 1 g/dL to 34 g/dL
Sensor condition, system failure, and low battery alarms
High Priority 571 Hz tone, 5-pulse burst, pulse spacing:
0.250s, 0.250s, 0.500s, 0.250s, repeat
time:10s
Medium Priority 550 Hz tone, 3-pulse burst, pulse spacing:
0.375s, 0.375s, repeat time: 7s
Low Priority 500 Hz tone, 1-pulse burst, repeat time: 5s
Alarm Muted Reminder 500Hz tone, 2-pulse burst, pulse spacing
0.375s, repeat time: 3 min
Alarm Volume High Priority: 70 dB (min)
Medium Priority: 70 dB (min)
Low Priority: 45 dB (min)