Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 107 Masimo
Low Signal IQ Message
Message: (Pulse CO-Ox) Low Signal IQ
Low signal quality.
Next steps: Ensure proper sensor application. Move sensor to a better perfused site. See
Signal IQ (SIQ) on page 111.
Low SpCO SIQ Message
SpCO measurement reading is obscured.
Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If
not, replace the sensor. See Successful Monitoring for SpCO on page 28.
Low SpMet SIQ Message
SpMet measurement reading is obscured.
Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If
not, replace the sensor. See Successful Monitoring for SpCO on page 28.
Low SpHb SIQ Message
SpHb measurement reading is obscured.
Next steps: Ensure proper sensor application. Check sensor to see if it is working properly. If
not, replace the sensor. See Successful Monitoring for SpHb on page 27.
Speaker Failure Message
Instrument requires service. Next steps: Contact Masimo Tech Support. Chapter 9: Service
and Maintenance on page 131.
No Cable Connected Message
Message:
(Pulse CO-Ox) No Cable Connected, or
(RAM) No Cable Connected
Cable not attached or not fully inserted into the connector.
Next steps: Disconnect and reconnect cable into connector.