Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 106 Masimo
No Adhesive Sensor Connected Message
Message:
(Pulse CO-Ox) No Adhesive Sensor Connected, or
(RAM) No Adhesive Sensor Connected
When a single-patient-use sensor is used, the adhesive portion of the sensor is not
connected. (Applies to Rainbow ReSposable CO-Oximeter Sensor Systems or ReSposable
Pulse CO-Oximeter Sensor Systems only.)
Next steps: Ensure the adhesive portion is firmly connected to the sensor.
Interference Detected Message
Message:
(Pulse CO-Ox) Interference Detected, or
(RAM) Interference Detected
High intensity light such as pulsating strobe lights, excessive ambient light sources such as
surgical lights or direct sunlight, or other monitor displays.
Next steps: Place a Masimo Optical Light Shield over the sensor.
Incorrect monitor line frequency setting (Hz).
Next steps: . Adjust the Line Frequency to the correct Hz setting. See Device Settings on
page 75.
SpO2 Only Mode Message
Message: (Pulse CO-Ox) SpO2 Only Mode
Occurs during an unsuccessful sensor initialization/pulse search routine or during
monitoring.
Next steps: See the directions for use provided with your sensor. Use a Masimo light shield to
cover the sensor and adjust the sensor.
Low Battery Message
Battery charge is low.
Next steps: Charge battery by placing the Handheld into the Docking Station and powering
the instrument with AC line power. Replace battery if necessary.
Low Perfusion Index Message
Message: (Pulse CO-Ox) Low Perfusion Index
Signal too small.
Next steps: Move sensor to better perfused site. See Low Perfusion on page 112.