Radical-7 Chapter 6: Alarms and Messages
www.masimo.com 104 Masimo
PI Delta Alarms
From the Alarms screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable Settings
Set Baseline See Perfusion Index (PI)
Delta Alarm Overview on
page 103.
Off On or Off
Baseline See Perfusion Index (PI)
Delta Alarm Overview on
page 103.
Off, or PI
baseline
N/A
Percent
Change
See Perfusion Index (PI)
Delta Alarm Overview on
page 103.
50% 10% to 99% in steps of 1%
Timeout See Perfusion Index (PI)
Delta Alarm Overview on
page 103.
None 1 min, 5 min, 30 min, 1 hr, 4
hr, 8 hr, 12 hr, 24 hr, 36 hr,
48 hr, or None.
Messages
The following section lists common messages, their potential causes, and next steps.
Replace Sensor Message
Message:
(Pulse CO-Ox) Replace Sensor, or
(RAM) Replace Sensor
SpHb reusable sensor has used all its available monitoring time. Sensor is non-functional.
Defective sensor.
Next steps: Replace sensor.