Service manual

Rad-8 Signal Extraction Pulse Oximeter Service Manual
4-4
4.9.2 Validation of the port can be done by connecting a serial printer
or using a PC with Hyper Terminal and the following setup:
4.9.3 Communication Parameters
baud rate 9600 bps
data bits 8
stop bit 1
parity none
Flow Control none
4.9.4 With the Rad-8 Pulse Oximeter connected as per the above
setup, attach a patient cable and sensor to the Rad-8 Pulse
Oximeter. Refer to page 4-11 of the Operator’s Manual and set
the output to ASCII 1. The printer (or PC monitor) will print in the
following format every second.
05/15/07 08:12:21 SPO2=---% BPM=--- ALARM=0000 EXC=000000