Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 95 Masimo
Sounds
Use the Sounds screen to control the volume of sounds and duration of audio pause on
Rad-97. Users can also access the Sounds screen by pressing the Sounds icon on the Status
Bar. See About the Status Bar on page 52.
Option Description Factory
Default
Settings
User Configurable
Settings
Alarm
Volume
Sets the alarm volume level.
4 (Highest
volume)
1 to 4 - Slide towards the
left to decrease volume
and to silence.
Pulse Tone
Volume
Sets the pulse tone volume level. 3
0 to 4 - Slide towards the
left to decrease volume
and to silence.
Audio Pause
Duration
Sets the length of time that the
audible alarm remains silenced,
when Audio Pause is enabled. See
Audio Pause on page 157.
2 minutes
1, 2, or 3 minutes,
Permanent *,**, or
Permanent with
Reminder *,***.
SmartTone Allows the audible pulse to continue
to beep when the pleth graph shows
signs of motion.
Off On or Off
* Requires user to have All Mute Enabled turned on in the Access Control menu. See Access
Control on page 107.
** If Permanent is selected, there will be no audible alarms, but visual alarms will still display.
*** If Permanent with Reminder is selected, a tone will sound every three (3) minutes as a
reminder that Permanent is active.