Rad-97 Chapter 4: Operation
www.masimo.com 94 Masimo
Options Description Factory Default
Settings
User Configurable Settings
RR 35 1 to 150 in steps of 1
Y-Axis Min
EtCO
2
,
FiCO
2
The Trend Min, indicating
the lowest value that will
be shown.
0.0 vol%
0.0 kPa
0 mmHg
0.0 to 24.9 vol% in steps of 0.1
0.0 to 29.9 kPa in steps of 0.1
0 to 224 mmHg in steps of 1
RR 0 0 to 149 in steps of 1
Additional Settings
Use the Additional Settings screen to configure the following:
Options Description Factory Default
Settings
User Configurable Settings
Apnea
Timeout
Set no-breath
timeout
30 seconds 20-60 seconds in 1 second intervals
Set O
2
Range Set O
2
range 0-30 vol% 0-30 vol%, 30-70 vol%, or 70-100 vol%
Set N
2
O
Range
Set N
2
O range 0-30 vol% 0-30 vol% or 30-70 vol%
CO
2
Unit of
Measure
Set CO
2
display
unit
mmHg
mmHg, kPa, or vol%
CAUTION: Selecting kPa will cause all
stored Trend data to be erased.