Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 93 Masimo
RR Alarms
From the Respiratory Rate screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit
High Limit is the upper
threshold that triggers an
alarm.
High 50
2 to 149 in steps of
1, or Off
When set to Off,
alarm is disabled
Low Limit
Low Limit is the lower threshold
that triggers an alarm.
High 6
Off, or 1 to 148 in
steps of 1
When set to Off,
alarm is disabled
Silence
Duration
Sets the amount of time that
the alarm is silenced.
NA 2 minutes
30 seconds, 1, 2, or
5 minutes
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
NA 30 seconds
0, 5, 10, 15, 20, 30,
or 60 seconds
Trends for NomoLine Capnography
From any of the capnography parameter setting screens, touch Trends, and then change any
of the following options:
Options Description Factory Default
Settings
User Configurable Settings
Y-Axis Max
EtCO
2
,
FiCO
2
The Trend Max, indicating
the highest value that will
be shown.
8.0 vol%
8.0 kPa
60 mmHg
0.1 to 25.0 vol% in steps of 0.1
0.1 to 30.0 kPa in steps of 0.1
1 to 225 mmHg in steps of 1