Rad-97 Chapter 4: Operation
www.masimo.com 93 Masimo
RR Alarms
From the Respiratory Rate screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit
High Limit is the upper
threshold that triggers an
alarm.
High 50
2 to 149 in steps of
1, or Off
When set to Off,
alarm is disabled
Low Limit
Low Limit is the lower threshold
that triggers an alarm.
High 6
Off, or 1 to 148 in
steps of 1
When set to Off,
alarm is disabled
Silence
Duration
Sets the amount of time that
the alarm is silenced.
NA 2 minutes
30 seconds, 1, 2, or
5 minutes
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
NA 30 seconds
0, 5, 10, 15, 20, 30,
or 60 seconds
Trends for NomoLine Capnography
From any of the capnography parameter setting screens, touch Trends, and then change any
of the following options:
Options Description Factory Default
Settings
User Configurable Settings
Y-Axis Max
EtCO
2
,
FiCO
2
The Trend Max, indicating
the highest value that will
be shown.
8.0 vol%
8.0 kPa
60 mmHg
0.1 to 25.0 vol% in steps of 0.1
0.1 to 30.0 kPa in steps of 0.1
1 to 225 mmHg in steps of 1