Rad-97 Chapter 4: Operation
www.masimo.com 92 Masimo
FiCO2 Alarms
From the Inspired Carbon Dioxide screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium
0.8 vol%
0.8 kPa
6 mmHg
0.2 to 24.9 vol% in
steps of 0.1, or Off
0.2 to 29.9 kPa in
steps of 0.1, or Off
2-224 mmHg in
steps of 1, or Off
When set to Off,
alarm is disabled
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
Low Off
Off, or 0.1 to 24.8
vol% in steps of 0.1
Off, or 0.1 to 29.8
kPa in steps of 0.1
Off, or 1-223 mmHg
in steps of 1
When set to Off,
alarm is disabled
Silence
Duration
Sets the amount of time that
the alarm is silenced.
NA 2 minutes
30 seconds, 1, 2, or
5 minutes
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
NA 30 seconds
0, 5, 10, 15, 20, 30,
or 60 seconds
RR Settings
From the Respiratory Rate Settings screen, access the following screens:
RR Alarms on page 93
About Parameter Information on page 83
Trends for NomoLine Capnography on page 93