Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 88 Masimo
Pulse Rate Alarms
From the Pulse Rate Settings screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium 120
40-215 in steps of 5,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium 50
Off, or 35-210 in
steps of 5
When set to Off,
alarm is disabled
Intervals Settings for NIBP
From the Intervals setting screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable
Settings
Set Mode The mode of measurement for NIBP. Automatic Automatic or Stat
Interval
Note: Option available when
Automatic mode is selected.
Automatic interval measurement mode
will take blood pressure measurements
once every desired interval.
15 minutes
2, 3, 4, 5, 10, 15, 30,
60, or 90 minutes
Stat
Duration
Note: Option available when Stat
mode
is selected.
Stat interval measurement mode will
take blood pressure measurements
continually for the desired duration.
10 minutes 5 or 10 minutes