Rad-97 Chapter 4: Operation
www.masimo.com 88 Masimo
Pulse Rate Alarms
From the Pulse Rate Settings screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium 120
40-215 in steps of 5,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium 50
Off, or 35-210 in
steps of 5
When set to Off,
alarm is disabled
Intervals Settings for NIBP
From the Intervals setting screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable
Settings
Set Mode The mode of measurement for NIBP. Automatic Automatic or Stat
Interval
Note: Option available when
Automatic mode is selected.
Automatic interval measurement mode
will take blood pressure measurements
once every desired interval.
15 minutes
2, 3, 4, 5, 10, 15, 30,
60, or 90 minutes
Stat
Duration
Note: Option available when Stat
mode
is selected.
Stat interval measurement mode will
take blood pressure measurements
continually for the desired duration.
10 minutes 5 or 10 minutes