Operator's Manual

Table Of Contents
Rad-97 Chapter 4: Operation
www.masimo.com 84 Masimo
Options Description Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium Off 80.2 °F to 109.9 °F, in
increments of 0.1, or Off
26.9 °C to 43.2 °C, in
increments of 0.1, or Off
When set to Off, alarms
are disabled.
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium Off 80.1 °F to 109.8 °F in
increments of 0.1, or Off
26.8 °C to 43.1 °C, in
increments of 0.1, or Off
When set to Off, alarms
are disabled.
Additional Settings
From the Additional Settings screen, change the following option:
Options Description Factory Default
Settings
User Configurable
Settings
Unit of Measure The unit of measure for
temperature.
°F °F, °C
Measurement
Timeout
Spot Check timing
customization.
5 minutes
5, 10, 15, 30, 60 and 90
minutes
Noninvasive Blood Pressure (NIBP) Settings
Note: This feature is optional on Rad-97 devices.
The NIBP menu allows the user to view and customize settings for the NIBP module by
changing any of the following options: