Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 84 Masimo
Options Description Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium Off 80.2 °F to 109.9 °F, in
increments of 0.1, or Off
26.9 °C to 43.2 °C, in
increments of 0.1, or Off
When set to Off, alarms
are disabled.
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium Off 80.1 °F to 109.8 °F in
increments of 0.1, or Off
26.8 °C to 43.1 °C, in
increments of 0.1, or Off
When set to Off, alarms
are disabled.
Additional Settings
From the Additional Settings screen, change the following option:
Options Description Factory Default
Settings
User Configurable
Settings
Unit of Measure The unit of measure for
temperature.
°F °F, °C
Measurement
Timeout
Spot Check timing
customization.
5 minutes
5, 10, 15, 30, 60 and 90
minutes
Noninvasive Blood Pressure (NIBP) Settings
Note: This feature is optional on Rad-97 devices.
The NIBP menu allows the user to view and customize settings for the NIBP module by
changing any of the following options: