Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 82 Masimo
Option Description Factory Default
Settings
User Configurable Settings
Waveform
Mode
Change the Waveform View.
See Waveform Mode on
page 82.
Pleth + Sig IQ +
Acoustic
Acoustic, Pleth + Sig IQ, Pleth
+ Sig IQ + Acoustic, PVi Pleth
+ Sig IQ, or PVi Pleth + Sig IQ
+ Acoustic
SmartTone Enable or disable the
SmartTone.
See Sounds on page 95.
Off On, Off
SpO
2
low %
limit
Set the SpO
2
low limit
alarm.
See SpO2 Settings on page
66.
Off Off, 1% to 98%
Waveform Mode
The following section contains examples of some of the waveforms viewable on the Main
Screen.
Signal IQ Indicators
Signal IQ (SIQ) indicators are displayed as vertical bars for each individual pulsation. The
height of the bar provides an assessment of the confidence in the SpO
2
measurement
displayed.
Acoustic Waveform View
The RRa waveform is located above the parameter values. Acoustic Respiratory Rate (RRa)
must be available for this feature to be shown. This view contains RRa waveform only.