Operator's Manual

Table Of Contents
Rad-97 Chapter 4: Operation
www.masimo.com 81 Masimo
Trends on page 110
About Parameter Information on page 83.
ORi Alarms
From the Alarms screen, change any of the following options:
Options Description Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit High Limit is the upper
threshold that triggers an
alarm.
Medium Off 0.02 to 0.99 or Off
in increments of
0.01
Low Limit Low Limit is the lower
threshold that triggers an
alarm.
Medium Off Off, or 0.01 to 0.98
in increments of
0.01
Trending
Down Alarm
Trending Down Alarm is
displayed when a rapid
decrease in ORi is measured.
Medium Off On or Off
Silence
Duration
Sets the amount of time that
the alarm is silenced.
N/A 2 minutes 30 seconds or 1, 2, or
5 minutes
Additional Settings
Use the Additional Settings screen to configure the following:
Option Description Factory Default
Settings
User Configurable Settings
Sensitivity
Mode
Change Sensitivity Mode.
See Sensitivity Modes
Overview on page 56.
APOD MAX, APOD, NORM