Rad-97 Chapter 4: Operation
www.masimo.com 77 Masimo
Additional Settings for SpHb
From the Additional Settings screen, change any of the following options:
Options Description Factory
Default
Settings
User
Configurable
Settings
Averaging Time The length of time over which the system
calculates the average of all data points.
Medium Short, Medium,
or Long
Arterial/Venous
Mode
Provides an arterial or venous value that
displays on the main screen.
Arterial Arterial or
Venous
Precision
(units of g/dL and
mmol/L)
Allows the user to set the precision of the
displayed SpHb value.
Note: When unit is g/L, Precision is
always 1 (whole numbers)
0.1 0.1, 0.5, or 1.0
Unit of Measure*
Displays total hemoglobin (SpHb) as g/dL
(grams per deciliter), g/L (grams per liter),
or mmol/L (millimoles per liter). Unit of
Measure cannot be changed during active
monitoring.
g/dL g/dL, g/L, or
mmol/L
*Changing Unit of Measure will delete all prior trend data for all parameters.
In Vivo for SpHb
In Vivo is currently not available in the U.S.A. and territories relying on FDA market
clearance.
From the In Vivo screen, change any of the following options:
Options Description Factory Default
Settings
User Configurable Settings
Enabled See In Vivo Adjustment
Overview on page 65.
Off On or Off
Offset
Amount
See In Vivo Adjustment
Overview on page 65.
0.0 when turned
on
-3.0 g/dL to +3.0 g/dL in
steps of ± 0.1 g/dL