Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 77 Masimo
Additional Settings for SpHb
From the Additional Settings screen, change any of the following options:
Options Description Factory
Default
Settings
User
Configurable
Settings
Averaging Time The length of time over which the system
calculates the average of all data points.
Medium Short, Medium,
or Long
Arterial/Venous
Mode
Provides an arterial or venous value that
displays on the main screen.
Arterial Arterial or
Venous
Precision
(units of g/dL and
mmol/L)
Allows the user to set the precision of the
displayed SpHb value.
Note: When unit is g/L, Precision is
always 1 (whole numbers)
0.1 0.1, 0.5, or 1.0
Unit of Measure*
Displays total hemoglobin (SpHb) as g/dL
(grams per deciliter), g/L (grams per liter),
or mmol/L (millimoles per liter). Unit of
Measure cannot be changed during active
monitoring.
g/dL g/dL, g/L, or
mmol/L
*Changing Unit of Measure will delete all prior trend data for all parameters.
In Vivo for SpHb
In Vivo is currently not available in the U.S.A. and territories relying on FDA market
clearance.
From the In Vivo screen, change any of the following options:
Options Description Factory Default
Settings
User Configurable Settings
Enabled See In Vivo Adjustment
Overview on page 65.
Off On or Off
Offset
Amount
See In Vivo Adjustment
Overview on page 65.
0.0 when turned
on
-3.0 g/dL to +3.0 g/dL in
steps of ± 0.1 g/dL