Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 4: Operation
www.masimo.com 69 Masimo
PR Alarms
From the PR Alarms screen, change any of the following options:
Options Description Alarm
Priority
Factory Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper threshold
that triggers an alarm.
High 140 bpm
35 bpm to 235
bpm, in steps of 5
bpm
Low
Limit
Low Limit is the lower threshold
that triggers an alarm.
High 50 bpm 30 bpm to 230
bpm, in steps of 5
bpm
Silence
Duration
Sets the amount of time that the
alarm is silenced.
NA 2 minutes 30 sec, 1, 2 or 5
minutes
Pi Settings
From the Pi Settings screen, access any of the following screens:
Pi Alarms on page 69
Additional Settings for Pi on page 70
Trends on page 110
About Parameter Information on page 83
Pi Delta on page 160
Pi Alarms
From the Alarms screen, change any of the following options:
Options Description Alarm
Priority
Factory
Default
Settings
User Configurable Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
Medium Off 0.04 to 0.09 in steps of 0.01
0.10 to 0.90 in steps of 0.10
1 to 19 in steps of 1, or Off