Operator's Manual

Table Of Contents
Rad-97 Chapter 4: Operation
www.masimo.com 65 Masimo
See SpO2 Settings on page 66
See PR Settings on page 68
See Pi Settings on page 69
See PVi Settings on page 70
See Respiration Rate (RR) Settings on page 71
See SpHb Settings on page 75
See SpOC Settings on page 78
See SpMet Settings on page 78
See SpCO Settings on page 79
See ORi Settings on page 80
In Vivo Adjustment Overview
In Vivo is currently not available in the U.S.A. and territories relying on FDA market
clearance.
The In Vivo Adjustment feature lets clinicians manually adjust the displayed value(s) of one
or more clinical parameters to match a corresponding laboratory reference during continuous
trending. To remind clinicians that the feature is active, an offset value displays alongside
the adjusted parameter value.
The In Vivo Adjustment feature for a parameter can be turned on by accessing the In Vivo
screen in the Settings menu for that parameter. After enabling the feature, set an offset
value. Once the feature is enabled, a positive or a negative offset value appears, as shown in
the below illustration.
The In Vivo offset amount is set to zero when any of the following conditions occur:
Cable or sensor is disconnected from device.
Sensor goes off patient causing a sensor initialization to occur.
Eight hours has elapsed since the In Vivo offset was activated.
Restore of factory defaults.
User turns off In Vivo.
Offset Value
When In Vivo Adjustment is activated for a specific parameter, the offset value appears
beneath that specific parameter. A positive value means that the displayed parameter value
has been increased (according to a laboratory reference value, as entered by a clinician) and a
negative value means the displayed parameter value has been decreased (according to a
laboratory reference value, as entered by a clinician).