Operator's Manual

Table Of Contents
Rad-97 Chapter 4: Operation
www.masimo.com 59 Masimo
Data can be added to or removed from Trend View in the same manner as described in
Customizing Windows on page 56. Data can be manipulated using the touchscreen as
follows:
1. Swipe the trend view display left or right to scroll the Trend View data forward or
backward in time.
2. Tap the Trend View in a specific spot to view the values at that time.
3. Touch the box in the lower right corner of the screen to change the time range of
Trend View data displayed on the screen. Select from 0:10h (10 minutes) to
24:00h (24 hours).
4. Using a pinch gesture with two fingers, zoom in and out of the Trend View data
displayed on the screen from 0:10h (10 minutes) to 24:00h (24 hours) in
increments of 0:01h.
About the System Status Light
The System Status Light provides visual indications of alarms and system messages. The
light will illuminate in different colors depending on the state of the device.
To locate the System Status Light, see Side and Top Views on page 41.
Light Status Alarm
Priority
Indication
None None System is off
Green None System is monitoring patient; no alarms.
Yellow Low
There is an active, low priority alarm.
Examples of low priority alarms:
No cable is connected.
Cable connected but there is no sensor connected to
cable.
Sensor is off patient and has been acknowledged.
Flashing
Yellow
Medium
There is an active medium priority alarm.
Example of a medium priority alarm:
Device battery is low.
Flashing Red High
There is an active high priority alarm.
Example of a high priority alarm:
The sensor has come off of the patient.