Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 2: Description
www.masimo.com 40 Masimo
Back View
1. Nurse Call Connector
Allows connection to a Nurse Call system.
CAUTION: To ensure patient electrical isolation, all
external device connections to the Analog
Output/Nurse Call connectors must be IEC 60950-1,
IEC 60601-1, or UL 1069 compliant.
See Nurse Call Connection on page 45.
2. Ethernet
Allows network connection to Rad-97 using an RJ-45
cable.
3. USB
Provides USB 2.0 connectivity.
4. Equipotential Ground Connector
Provides optional functional earthing for Rad-97 to
eliminate potential differences between the earth
connections for Rad-97 and another medical device.
The use of the Equipotential Ground Connector should
be in accordance with IEC 60601-1.
5. Power Entry Module
Provides connection to an AC power cord.
Note: Always connect the Rad-97 to the mains power
for continuous operation and/or battery recharging.
Note: To disconnect the device from AC power, first
disconnect the power cord from the power outlet,
rather than from the device.
6. Capnography Gas Sample Exhaust Port*
Exhaust port for gas samples.
*Optional feature. Only available on Rad-97 equipped
with NomoLine Capnography capabilities.