Rad-97 Chapter 2: Description
www.masimo.com 40 Masimo
Back View
1. Nurse Call Connector
Allows connection to a Nurse Call system.
CAUTION: To ensure patient electrical isolation, all
external device connections to the Analog
Output/Nurse Call connectors must be IEC 60950-1,
IEC 60601-1, or UL 1069 compliant.
See Nurse Call Connection on page 45.
2. Ethernet
Allows network connection to Rad-97 using an RJ-45
cable.
3. USB
Provides USB 2.0 connectivity.
4. Equipotential Ground Connector
Provides optional functional earthing for Rad-97 to
eliminate potential differences between the earth
connections for Rad-97 and another medical device.
The use of the Equipotential Ground Connector should
be in accordance with IEC 60601-1.
5. Power Entry Module
Provides connection to an AC power cord.
Note: Always connect the Rad-97 to the mains power
for continuous operation and/or battery recharging.
Note: To disconnect the device from AC power, first
disconnect the power cord from the power outlet,
rather than from the device.
6. Capnography Gas Sample Exhaust Port*
Exhaust port for gas samples.
*Optional feature. Only available on Rad-97 equipped
with NomoLine Capnography capabilities.