Operator's Manual

Table Of Contents
Rad-97 Chapter 16: Service and Maintenance
www.masimo.com 209 Masimo
8. If one of the overpressure values is not within the pass criteria above, return the
module for service to an authorized service center.
Maintenance
Battery Operation and Maintenance
The Rad-97 includes a lithium ion rechargeable battery.
Before using the Rad-97 without the AC power connected, check the battery status indicator
and ensure that the battery is fully charged. See Battery Charge Status Indicator on page 54.
To charge the Rad-97 battery, refer to Initial Battery Charging on page 44.
Note: When battery run time is significantly reduced, it is advisable to completely discharge
and fully recharge the battery.
Run Time for Rad-97
The following table outlines the estimated minimum run time of the battery in the Rad-97.
The time estimates are based on a fully charged battery.
Time estimates are also based on specific operating modes.
For optimal battery run time, configure the device to automatically adjust the brightness. See
Brightness on page 107.
Configuration Operating Mode Minimum run time (Est.)
Rad-97
Not connected to AC power
Wireless connected
Bluetooth connected
Brightness level set to maximum
2 hours
NomoLine Capnography
Maintenance
Once every year, or whenever gas readings are questionable, perform a leakage check and
verify gas readings with a reference instrument or with calibration gas. See Leakage Check on
page 210.
Gas Bottle Disposal Procedures
For proper disposal of empty gas bottles, perform the following procedure: