Operator's Manual

Table Of Contents
Rad-97 Chapter 16: Service and Maintenance
www.masimo.com 208 Masimo
7. Calibration is completed.
Span Point Calibration
Note: This section is provided as a reference and intended for qualified service professionals
only.
Calibration Steps
1. Enter NIBP menu and select "Calibration".
2. Enter password: 4258 and select "Span Point Calibration".
3. Connect a manometer, volume and the hand bulb to the Module using "T" adapters
and connection tubing.
4. Apply 250 mmHg to the module.
5. Touch the "Calibrate" to start the span point calibration.
6. Results are displayed for Span Point Calibration.
7. Calibration is completed.
Overpressure Test
Note: This section is provided as a reference and intended for qualified service professionals
only.
Note: This test is performed from the Main Screen.
Pass Criteria
International standards for automated NIBP devices require that the pressure must not
exceed 300 mmHg on adults and pediatric patients and 150 mmHg on neonatal patients
with a tolerance of 10% for 15 seconds or greater than 10% for 3 seconds. The overpressure
pass criteria for the Advantage module are:
Adults, Pediatrics 300 ± 10mmHg
Neonates 150 ± 5mmHg
Test Method
The steps outlined below are for manually performing an overpressure test. Some or all of
these steps may be incorporated into a service tool provided by the medical device
manufacturer.
1. Set the corresponding patient profile. See Chapter 5: Profiles on page 117.
2. Perform an NiBP measurement.
3. Connect a manometer, volume and hand bulb to Module using "T" adapters.
4. From the Main Screen, touch the Start NIBP measurement button.
5. Touch the "Test" button on the display.
6. Increase the pressure to approximately 250mmHg using the hand bulb.
7. VERY SLOWLY increase the pressure from 280 to the overpressure point. Once it is
reached, the valves will open and the pressure will rapidly reduce to 0mmHg.