Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 16: Service and Maintenance
www.masimo.com 207 Masimo
Procedure
1. Enter NIBP menu, select "Calibration".
2. Enter password: 4258 and touch "Calibration Test".
3. Connect a manometer, volume and the hand bulb to the Module using "T" adapters
and connection tubing.
4. Touch the "Test" button on the display to start the calibration.
5. Apply various pressures (between 0 mmHg and 280 mmHg) to the Module with
the hand bulb. Verify that the Module pressure is equal to the manometer pressure
(±3mmHg). If the Module pressure does not agree with the manometer pressure
(±3mmHg), perform the Zero Point Calibration and the Span Point Calibration.
Then perform this calibration again.
6. Calibration is now complete.
NIBP Air Leak Test
Note: This section is provided as a reference and intended for qualified service professionals
only.
Pass Criteria
International standards for automated NIBP devices require that air leakage within the
pneumatic system must not exceed 6mmHg/minute.
Procedure
1. Connect the manometer and rigid volume (500 mL bottle) to the air hose
connection using "T" adapters and connection tubing.
2. Enter NIBP menu, select "Calibration".
3. Enter password: 4258 and select "Air Leak Test".
4. Touch the "Test" button on the display to start the test.
5. Wait for the countdown timer to reach 0 second.
6. Check the "Result" section, if the leak rate is greater than 6 mmHg/min, contact
customer service.
Zero Point Calibration
Note: This section is provided as a reference and intended for qualified service professionals
only.
Calibration Steps
1. Enter NIBP menu, select "Calibration".
2. Enter password: 4258 and select "Zero Point Calibration".
3. Connect a manometer, volume and the hand bulb to the Module using "T" adapters
and connection tubing.
4. Apply 0 (zero) mmHg to the module.
5. Touch the "Calibrate" to start the zero point calibration.
6. Results are displayed for Zero Point Calibration.