Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 15: Specifications
www.masimo.com 200 Masimo
Symbol Description Symbol Description
RS-232 Interface
Ethernet
Analog Out Interface
USB port
Greater than
China Restriction of
Hazardous Substances
Less than
The names and content of
the
toxic and hazardous
substances or elements shall
be provided in the product
instruction manual
Instructions/Directions for Use/Manuals are available in electronic format
@http://www.Masimo.com/TechDocs
Note: eIFU is not available in all countries.
Citations
[1] Refer to sensor DFU for specific sensor performance specifications.
[2] SpO2, SpCO, and SpMet accuracy was determined by testing on healthy adult volunteers in
the range 60% to 100% SpO2, 0% to 40% SpCO, and 0% to 15% SpMet against a laboratory
CO-Oximeter. SpO2 and SpMet accuracy was determined on 16 neonatal NICU patients ranging
in age from 7 days to 135 days old and weighing between 0.5 kg and 4.25 kg. Seventy-nine (79)
data samples were collected over a range of 70% to 100% SaO2 and 0.5% to 2.5% HbMet with a
resultant accuracy of 2.9% SpO2 and 0.9% SpMet. Contact Masimo for testing specifications.
[3] The Masimo rainbow SET technology with Masimo sensors has been validated for no motion
accuracy in human blood studies on healthy adult male and female volunteers with light to
dark skin pigmentation in induced hypoxia studies in the range of 70%-100% SpO
2
against a
laboratory CO-Oximeter and ECG monitor.
[4] The Masimo rainbow SET technology with Masimo sensors has been validated for motion
accuracy in human blood studies on healthy adult male and female volunteers with light to
dark skin pigmentation in induced hypoxia studies while performing rubbing and touching
motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5
Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70%-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor.
[5] The Rad-97 has been validated for low perfusion accuracy in bench-top testing against a
Biotek Index 2™* simulator and Masimo's simulator with signal strengths of greater than 0.02%
and transmission of greater than 5% for saturations ranging from 70%-100%.