Operator's Manual

Table Of Contents
Rad-97 Chapter 15: Specifications
www.masimo.com 187 Masimo
Effects from Water Vapor Partial Pressure on Gas Readings
When the breathing gas flows through the sampling line, the gas temperature will adapt to
the ambient temperature before reaching the gas analyzer. The measurement of all gases will
always show the actual partial pressure at the current humidity level in the gas sample. As
the NOMO section removes all condensed water, no water will reach the NomoLine
Capnography gas analyzer. However at an ambient temperature of 37 °C and a breathing gas
with a relative humidity of 95% the gas reading will typically be 6% lower than corresponding
partial pressure after removal of all water.
Electrical
AC Power Requirements
AC Power requirements 100 to 240 VAC, 47 to 63 Hz
Power consumption 60 VA
Fuses
UL, Metric (5x20mm), rated min. 250 VAC, 1 Amp, Time Delay,
min.1500A breaking capacity
Battery
Type Lithium ion
Capacity Approx. 4 hours [11]
Charging Time 6 hours [12]
Environmental
Environmental Conditions
Operating Temperature 0°C to 40°C (32°F to 104°F)
Storage/Transport Temperature -20°C to 60°C (-4°F to 140°F) [13]
Operating Humidity