Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 15: Specifications
www.masimo.com 186 Masimo
Gas or Vapor Gas Level CO
2
HAL
4
4 vol% _
1
ENF, ISO, SEV
4
5 vol% +8% of reading
3
DES
4
15 vol% +12% of reading
3
Xe (Xenon)
4
80 vol% -10% of reading
3
He (Helium)
4
50 vol% -6% of reading
3
Metered dose inhaler propellant
4
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
4
0.3 vol% _
1
C
3
H
7
OH (Isopropanol)
4
0.5 vol% _
1
CH
3
COCH
3
(Acetone)
4
1 vol% _
1
CH
4
(Methane)
4
3 vol% _
1
CO (Carbon Monoxide)
5
1 vol% _
1
NO (Nitrogen Monoxide)
5
0.02 vol% _
1
O
2
5
100 vol% _
2
Note 1: Negligible interference, effect included in the specification “Accuracy, all
conditions” above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included
in the specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically
decreases the CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0
vol% CO
2
and 50 vol% Helium, the actual measured CO
2
concentration will typically be
(1-0.06) * 5.0 vol% = 4.7 vol% CO
2
.
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.