Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 15: Specifications
www.masimo.com 185 Masimo
Data Output Specifications
Fi and ET** Fi and ET are displayed after one breath and have a continuously updated
breath average.
The following method is used to calculate end-tidal (ET) values:
The highest concentration of CO
2
during one breathing cycle with a weight
function applied to favor values closer to the end of the cycle.
ET will typically decrease below nominal value (ET
nom
) when respiration rate
(RR) exceeds the RR threshold (RR
th
) according to the following formula for
CO
2
:
ET=ET
nom
×(125/RR) for RR
th
>125
**Measured at I/E ratio 1:1 using breath simulator according to EN ISO 80601-2-55 fig.
201.101.
Gas Analyzer Specifications
Sensor head
• 2 channel NDIR type gas analyzer measuring at 4 to 5 μm.
• Data acquisition rate 10 kHz (sample rate 20 Hz / channel).
Compensations
• Automatic compensation for pressure and temperature.
• Manual compensation for broadening effects on CO
2
.
Calibration No span calibration is required for the IR bench. An automatic
zeroing is performed 1 to 3 times per day.
Warm-up time < 10 seconds (Concentrations reported and full accuracy)
Rise time*** ≤ 200 ms
Total System Response
Time
< 3 seconds (with 2 m NomoLine Airway Adapter Set sampling line)
***Measured according to EN ISO 80601-2-55.
Interfering Gas Vapor Effect
Gas or Vapor Gas Level CO
2
N
2
O
4
60 vol% _
2