Rad-97 Chapter 15: Specifications
www.masimo.com 184 Masimo
Standard Conditions
The following accuracy specifications are valid for dry single gasses at 22 ± 5 °C and 1013 ±
40 hPa:
Gas Range* Accuracy
CO
2
0 to 15 vol%
15 to 25 vol%
±(0.2 vol% + 2% of reading)
Unspecified
*All gas concentrations are reported in units of volume percent and may be translated into
mmHg or kPa by using the reported atmospheric pressure.
All Conditions
The following accuracy specifications are valid for all specified environmental conditions:
Gas Accuracy
CO
2
±(0.3 kPa + 4% of reading)
Additional NomoLine Capnography Specifications
General Specifications
Ambient CO
2
≤ 800 ppm (0.08 vol%)
Water Handling NomoLine Family sampling lines with proprietary water removal
tubing.
Sampling Flow Rate
50 ± 10 sml/min*
*Volumetric flow rate of air corrected to standardized conditions of temperature and pressure.
Data Output Specifications
Breath
detection
Adaptive threshold, minimum 1 vol% change in CO
2
concentration.
Respiration
rate**
0 to 150 ± 1 breaths/min