Rad-97 Chapter 15: Specifications
www.masimo.com 182 Masimo
Noninvasive Blood Pressure (NIBP) Specifications
Display Range
Measurement Patient Population Display Range
Systolic Adult 40-260 mmHg
Pediatric 40-230 mmHg
Neonatal 40-130 mmHg
Diastolic Adult 20-200 mmHg
Pediatric 20-160 mmHg
Neonatal 20-100 mmHg
MAP Adult 26-220 mmHg
Pediatric 26-183 mmHg
Neonatal 26-110 mmHg
Pulse Rate (PR) All 30-220 bpm
Accuracy
Pressure Transducer
Between 0 mmHg and 300 mmHg ±3 mmHg
Blood Pressure
Meets ANSI/AAMI SP10 and ISO 81060-2 (Mean difference of ≤5 mmHg with a standard
deviation of ≤8 mmHg)