Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 15: Specifications
www.masimo.com 180 Masimo
Measurement A
RMS
Values for Disposable (RD SET Series) Sensors
SpO
2
Accuracy Range (%) A
RMS
(%)
90-100 1.53
70-100 1.15
The table below provides the upper 95% and lower 95% limits of agreement. The differences
between measurements by the two methods are used to calculate the mean and standard
deviation. The lower 95% limit of agreement is the mean minus 1.96 standard deviation and
the upper 95% limit of agreement is the mean plus 1.96 standard deviation. These limits are
expected to contain 95% of the differences between measurements between the two methods
in controlled environments.
SpO
2
Upper and Lower Limits of Agreement (LoA)*
Upper 95% LoA 2.27%
Lower 95% LoA -2.29%
* See Bland and Altman. Agreement between methods of measurement with multiple
observations per individual. Journal of Biopharmaceutical Statistics (2007) vol. 17 pp.
571-582.
The below Bland-Altman plot represents the correlation of the (SpO
2
+ SaO
2
)/2 versus (SpO
2
-
SaO
2
) under no motion with an upper 95% and lower 95% limits of agreement.
Figure 1: Disposable (RD SET Series) Sensors (ARMS 70-100%)