Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 15: Specifications
www.masimo.com 178 Masimo
Accuracy (ARMS*) [1]
Oxygen Saturation (SpO
2
)
No Motion [2]
(SpO
2
from 60% to 80%)
Adults, Pediatrics, Infants 3%
No Motion [3]
(SpO
2
from 70% to 100%)
Adults, Pediatrics, Infants 2%
Neonates 3%
Motion [4]
(SpO
2
from 70% to 100%)
All patient populations
3%
Low perfusion [5]
(SpO
2
from 70% to 100%)
All patient populations
2%
Pulse Rate (PR)
Range 25 to 240 bpm
No motion All patient populations 3 bpm
Motion [6] All patient populations 5 bpm
Low Perfusion All patient populations 3 bpm
SpO
2
Upper and Lower Limits of Agreement (LoA)**
Upper 95% LoA [7] 2%
Lower 95% LoA [7] -2%
Carboxyhemoglobin Level (SpCO) [2]
Range of 1% to 40% Adults, Pediatrics, Infants 3%
Methemoglobin Level (SpMet) [2]
Range 1% to 15% All patient populations 1%