Operator's Manual

Table Of Contents
Rad-97 Safety Information, Warnings and Cautions
www.masimo.com 16 Masimo
instruments prior to clinical decision making to completely understand the patient’s
condition.
WARNING: Rad-97 is not an apnea monitor.
WARNING: Rad-97 should not be used as a replacement or substitute for ECG-based
arrhythmia analysis.
WARNING: Rad-97 may be used during defibrillation. This may affect the accuracy or
availability of the parameters and measurements.
WARNING: Rad-97 may be used during electrocautery. This may affect the accuracy or
availability of the parameters and measurements.
WARNING: Always check that speaker is functional prior to use.
WARNING: Avoid placing Rad-97 against a surface that may cause the alarm to be muffled.
This may result in the inability to detect the audible alarms.
WARNING: Rad-97 may not fully charge in a high ambient temperature environment.
WARNING: Properly apply sensors according to sensor's directions for use. Misapplied sensor
or sensors that become partially dislodged may cause no or incorrect readings.
WARNING: Select a well perfused site for monitoring, very low perfusion at the monitored site
may result in no or incorrect readings.
WARNING: Do not use Rad-97 on patients that have been injected with dyes or any
substance containing dyes, the change usual blood pigmentation may cause no or incorrect
readings.
WARNING: Display parameter may not be accurate when a low SIQ message is provided.
Clinicians should consider additional information to supplement values to completely
understand the patient’s condition.
WARNING: If SpO
2
values indicate hypoxemia, a laboratory blood sample should be taken to
confirm the patient’s condition.
WARNING: SpO
2
is empirically calibrated in healthy adult volunteers with normal levels of
carboxyhemoglobin (COHb) and methemoglobin (MetHb).
WARNING: Optical, pleth-based measurements (e.g. SpO
2
, SpHb, SpOC, SpMet, SpCO, RRp,
and ORi) can be affected by the following:
Improper sensor application or use of incorrect sensor.
Blood pressure cuff applied to the same arm as the sensor site.
Intravascular dyes such as indocyanine green or methylene blue.
Venous congestion.
Abnormal venous pulsations (e.g. tricuspid value regurgitation, Trendelenburg
position).
Abnormal pulse rhythms due to physiological conditions or induced through
external factors (e.g. cardiac arrhythmias, intra-aortic balloon, etc.).
Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc.
Moisture, birthmarks, skin discoloration, nail aberration, deformed fingers, or
foreign objects in the light path.
Elevated levels of bilirubin.