Rad-97 Chapter 14: Troubleshooting
www.masimo.com 171 Masimo
Symptom Potential Causes Next Steps
Dimly Lit Parameters
Low signal quality. Assess the patient.
Verify sensor type and size and
re-apply sensor. See Directions for
Use for sensor.
Check if blood flow to the sensor
site is restricted.
Check the placement of the
sensor. Re-apply sensor or move
to a different site.
Replace sensor.
Minimize or eliminate motion at
the monitoring site.
Set to MAX Sensitivity. See
Sensitivity Modes Overview on
page 56.
Parameter Values Do Not
Correlate With Clinical
Assessment or Arterial
Blood Gas
Measurements
Low perfusion.
Sensor
displacement.
Check for error messages. See
Chapter 13: Alarms and Messages
on page 153.
Check placement of sensor or if it
is too tight. Reapply sensor or
select a new site. Set to MAX
sensitivity and confirm that the
sensor is securely placed on the
patient. See Directions for Use for
sensor.
Unexpected Parameter
Readings
Low SIQ or Pi
values.
Inappropriate
sensor size or
sensor
measurement
location.
Reposition sensor to site with
strong SIQ and Pi. Average
readings taken from three
different sites to improve
accuracy. Submit blood sample for
laboratory CO-Oximetry test for
comparison.
Verify proper sensor for patient
size. Verify proper sensor site. See
Directions for Use for sensor.