Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 14: Troubleshooting
www.masimo.com 170 Masimo
Symptom Potential Causes Next Steps
Difficulty obtaining a
reading.
• Inappropriate
sensor or sensor
size.
• Improper sensor
type or application.
• Low perfusion.
• Excessive motion
artifact.
• Excessive ambient
or strobing light.
• Low battery/ not
plugged into AC
power supply.
• Interference from
line
frequency-induced
noise.
• Allow time for parameter reading
to stabilize.
• Verify sensor type and size and
re-apply sensor. See Directions for
Use for sensor.
• Check if blood flow to the sensor
site is restricted.
• Check the placement of the
sensor. Re-apply sensor or move
to a different site.
• Replace sensor.
• Verify the device and sensor are
configured with the parameter.
• Verify proper sensor and sensor
size for the patient.
• Shield the sensor from excessive
or strobing light.
• Minimize or eliminate motion at
the monitoring site.
• Connect AC power supply.
• Verify and set 50 or 60Hz menu
setting. See Localization on page
97.
Parameter readings
displayed as dashes.
• Parameter may not
have stabilized.
• Device may not be
configured with the
parameter.
• Sensor is not
compatible with
the parameter.
• Allow time for parameter reading
to stabilize.
• Verify sensor type and size and
re-apply sensor. See Directions for
Use for sensor.
• Check if blood flow to the sensor
site is restricted.
• Check the placement of the
sensor. Re-apply sensor or move
to a different site.
• Replace sensor.
• Verify the device and sensor are
configured with the parameter.