Rad-97 Chapter 13: Alarms and Messages
www.masimo.com 166 Masimo
NomoLine Capnography Messages
The following section lists common NomoLine capnography messages, their potential causes,
and next steps.
Message Potential Causes Next Steps
No Breath
Detected
Indicates no breath is
detected within selected
apnea timeout setting.
Confirm patient is properly
connected to sampling line.
Check patient breathing circuit.
Replace sampling line.
Replace
Sampling Line
Indicates that a sampling
line is clogged or blocked,
not allowing flow.
Indicates that the
sampling line should be
replaced.
Check for a blockage and replace
the sampling line if necessary.
Replace the sampling line
No Sampling
Line
Indicates that a sampling
line is not connected to
Rad-97.
Reconnect the sampling line
Check Ambient
Temperature
Indicates that the internal
temperature is out of
range.
Confirm standard operating
conditions.
Contact Masimo Technical
Support. See Chapter 16: Service
and Maintenance on page 203.
Check Ambient
Pressure
Indicates ambient
pressure measurement
out of range.
Confirm standard operating
conditions.
Contact Masimo Technical
Support. See Chapter 16: Service
and Maintenance on page 203.
Replace Module
Indicates sensor error.
Indicates hardware error.
Indicates software error.
Contact Masimo Technical
Support.
See Chapter 16: Service and
Maintenance on page 203.
Low EtCO
2
SIQ
Low FiCO
2
SIQ
EtCO
2
or FiCO
2
measurement is outside
specified accuracy range.
If the message is persistent, verify
gas reading with reference
instrument or with calibration gas.
Ensure patient is properly
connected to sampling line.
Zeroing Required
Indicates zeroing to
environmental
Perform a leakage check. See
Leakage Check on page 210.