Operator's Manual

Table Of Contents
Rad-97 Chapter 13: Alarms and Messages
www.masimo.com 165 Masimo
Noninvasive Blood Pressure (NIBP) Messages
The following section lists common NIBP messages, their potential causes, and next steps.
Message Potential Causes Next Steps
Check Cuff (Weak
Signal)
Weak or no signal measured
during blood pressure
measurement.
Check that the hose is
connected.
Check that the correct size
cuff is being applied.
Check that the cuff is in
the correct position.
Check that there is no
excessive clothing
between arm and cuff.
Retake another
measurement.
Check that the cuff is not
leaking air.
If problem still persists,
contact Customer Service.
Check Cuff
(Artifact)
Motion may be affecting ability
to take measurement.
Check Cuff (Out of
Range)
Measurement is out of range.
Check Cuff
(Measurement
Timeout)
Weak signal when measurement
is being taken.
Check Cuff
(Pneumatic
Blockage)
May be a blockage in the air
hose.
Check Cuff (Inflate
Timeout)
May be a blockage in the air
hose.
Check Cuff (Safety
Timeout)
Weak signal when measurement
is being taken.
Check Cuff
(Overpressure)
May be due to a faulty cuff.
Calibration
Required
Blood pressure measurement
transducer may be out of range
or there has been a calibration
data failure.
Perform calibration
procedures. See Chapter
16: Service and
Maintenance on page
203.
If problem still persists,
contact Masimo Technical
Support.
Module Error
Device requires service.
Contact Masimo Technical
Support. See Chapter 16:
Service and Maintenance
on page 203.