Operator's Manual
Table Of Contents
- About This Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1: Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- Pulse CO-Oximetry vs. Drawn Whole Blood Measurements
- General Description for Total Hemoglobin (SpHb)
- General Description for Total Arterial Oxygen Content (CaO2)
- General Description for SpOC
- General Description for Carboxyhemoglobin (SpCO)
- General Description for Methemoglobin (SpMet)
- General Description for Respiration Rate (RRp)
- General Description for Oxygen Reserve Index (ORi)
- SpCO, SpMet, and SpHb Measurements During Patient Motion
- rainbow Acoustic Monitoring™ (RAM™)
- Chapter 2: Description
- Chapter 3: Setting Up
- Chapter 4: Operation
- Using the Touchscreen and Home Button
- About the Main Screen
- About the System Status Light
- Accessing Main Menu Options
- rainbow Parameter Settings
- Parameter Settings
- Temperature Settings
- Noninvasive Blood Pressure (NIBP) Settings
- NomoLine Capnography Settings
- Sounds
- Device Settings
- About
- Trends
- Call
- Rad-97 Screenshot Capture
- Patient Admit/Discharge
- EMR Push
- Chapter 5: Profiles
- Chapter 6: Temperature
- Chapter 7: Noninvasive Blood Pressure (NIBP)
- Chapter 8: NomoLine Capnography
- Chapter 9: Video Conferencing
- Chapter 10: Admit to and Discharge from Patient SafetyNet
- Chapter 11: Electronic Medical Records (EMR) Push
- Chapter 12: Third-Party Devices
- Chapter 13: Alarms and Messages
- Chapter 14: Troubleshooting
- Chapter 15: Specifications
- Pulse CO-Oximetry Specifications
- Temperature Specifications
- Noninvasive Blood Pressure (NIBP) Specifications
- NomoLine Capnography Specifications
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- Compliance
- Connectors
- Wireless Specifications
- Guidance and Manufacturer's Declaration-Electromagnetic Emissions
- Guidance and Manufacturer's Declaration-Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 16: Service and Maintenance
- Appendix: Concepts of Alarm Response Delay
- Index
Rad-97 Chapter 13: Alarms and Messages
www.masimo.com 160 Masimo
*De Felice C, Latini G, Vacca P, Kopotic RJ. The pulse oximeter perfusion index as a predictor for
high illness severity in neonates. Eur J Pediatr. 2002;161:561-562.
Pi Delta
From the Pi Delta menu screen, change any of the following options:
Options Description Factory
Default
Settings
User Configurable
Settings
Set
Baseline
Sets the Perfusion Index (Pi) value to be
used as the baseline.
Off On or Off
Percent
Change
The change in Pi from the baseline that,
if maintained for the Timeout length,
will trigger audible and visual alarms.
50% 10% to 99%, in steps of
1%
Timeout The length of time over which the
percent change in Pi is monitored.
None None or 1, 5, 30
minutes, 1, 4, 8, 12, 24,
36, 48 hours
Rad-97 Messages
The following section lists common messages, their potential causes, and next steps.
Message Potential Causes Next Steps
(Pulse CO-Ox)
Replace Cable
or
(RAM) Replace Cable
• The patient cable is
non-functional or the patient
monitoring time of the cable
has expired.
• Replace the patient cable.
(Pulse CO-Ox) Cable
Near Expiration
or
(RAM) Cable Near
Expiration
• Patient cable has less than
10% of patient monitoring
time remaining.
• Replace with new patient
cable.