Rad-97 Chapter 8: NomoLine Capnography
www.masimo.com 133 Masimo
Two different ways of scavenging the exhaust gas; through a ventilator reservoir
(Figure 1) and directly to a scavenging system (Figure 2). In both cases, generate a
deliberate leakage to avoid vacuum in the ISA’s exhaust gas tube.
Note: Returning the NomoLine capnography exhaust gas to the patient circuit is
not allowed in the USA (Figure 3).
3. Power up Rad-97. See Powering the Rad-97 ON and OFF on page 44.
4. The green LEGI Indicator shows that the NomoLine Capnography analyzer is ready
for use. See NomoLine Capnography LEGI Indicator on page 60.
Note: Without a NomoLine sampling line connected, the LEGI Indicator will not
illuminate.
5. Perform a pre-use check as described in Pre-Use Check on page 137.
6. Attach NomoLine to the patient for monitoring. Refer to the NomoLine Directions
for Use.
7. Parameters will display in the Capnography window. See Capnography Display on
page 134.