Rad-97 Chapter 7: Noninvasive Blood Pressure (NIBP)
www.masimo.com 128 Masimo
6. The device will begin the Automatic Interval Measurement and display the
specified interval designated by the clinician (15 minutes in the example below).
7. Wait for measurement to complete or touch the Stop button to stop the
measurement.
Note: Once finished measuring, values will appear and the next measurement will
begin after the specified interval.
Stat Interval NIBP Measurement
Stat Interval Measurement mode will take blood pressure measurements continuously for the
desired duration.
To measure blood pressure in Stat Interval Measurement Mode
1. Ensure that the correct patient profile is selected before measurement.
WARNING: Only use Rad-97 in Neonatal mode with a neonatal blood pressure cuff
to measure blood pressure on neonates.
2. To change the patient profile, touch the Main Menu icon , then select Profiles
.
Touch the Profile Name to select the desired patient profile.
3. Properly place the blood pressure cuff on patient. See Cuff Selection and
Placement on page 125.