Rad-97 Chapter 4: Operation
www.masimo.com 108 Masimo
Option Description Factory
Default
Settings
User Configurable Settings
Power on
Profile
Sets the profile used when the
device is powered on. See
Chapter 5: Profiles on page 117.
Previous
Profile
Previous Profile, Adult,
Pediatric, Neonatal, or User
defined profile (up to 8)
All Mute
Enabled
Enables parameter Alarm Silence
menu option. See Sounds on
page 95.
Off On or Off
Lock Alarm
Volume
Sets the lowest alarm volume
level.
Off 3, 4, or Off
Screen Lock
Allows the user to lock the
touchscreen to prevent
accidental changes.
Off On or Off
USB Port
Baudrate
Sets the USB port
communication speed.
921600
9600, 19200, 38400,
57600, 115200, 230400,
or 921600
Data
Collection
Enabled
Enables or disables physical data
collection mode.
Off On or Off
Save as
Adult*
Saves current profile parameter
as the Adult Profile.
N/A
Press Save to update the
profile.
Save as
Pediatric*
Saves current profile parameter
as the Pediatric Profile.
N/A
Press Save to update the
profile.
Save as Neo*
Saves current profile parameter
as the Neonatal Profile.
N/A
Press Save to update the
profile.
Factory
Defaults
Options are restored to factory
values.
N/A Press Restore.
* See Replacing Factory Default Settings for Adult, Pediatric and Neonatal Profiles on page
119.