Operator's Manual

Root Chapter 3: Operation
www.masimo.com 61 Masimo
Specific ranges of the Analog View are:
Ref.
Feature
Color
Description
1
Normal
Range
White
Area of the display range where an alarm will
not be triggered.
2
Caution
Range
Yellow
Area of the display range that provides a
caution indicator.
3
Alarming
Range
Red
Area of the display range where an alarm will
be triggered.
Some ranges display as quarter circles, others display as half circles. A
quarter circle displays when the value has a physiologic normal level at one
end of the range. A half circle displays when the value has a physiologic
normal level in the middle of the display range.
In the example below, the SpO
2
gauge is shown as a quarter circle, where
values lower than 88% will trigger an alarm, and the PR gauge is shown as
a half circle, where values below 50 bpm and above 140 bpm will trigger an
alarm.