Operator's Manual

Root Safety Information, Warnings, and Cautions
www.masimo.com 23 Masimo
CAUTION: Do not touch, press, or rub the display panels with abrasive
cleaning compounds, instruments, brushes, rough-surface materials, or
bring them into contact with anything that could scratch the display.
CAUTION: Do not submerge the Root in any cleaning solution or attempt
to sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other
method. This will seriously damage the device.
CAUTION: Electrical shock and flammability hazard: Before cleaning, always
turn off the device and disconnect from any AC power source.
CAUTION: An operator may only perform maintenance procedures
specifically described in the manual. Refer servicing to qualified service
personnel trained in the repair of this equipment.
CAUTION: Electrical Shock Hazard: Carry out periodic tests to verify that
leakage currents of patient-applied circuits and the system are within
acceptable limits as specified by the applicable safety standards. The
summation of leakage currents must be checked and in compliance with IEC
60601-1 and UL60601-1. The system leakage current must be checked when
connecting external equipment to the system. When an event such as a
component drop of approximately 1 meter or greater or a spillage of blood
or other liquids occurs, retest before further use. Injury to personnel could
occur.
Note: Excessive cleaning solution can flow into the device and cause
damage to internal components.
Compliance Warnings and Cautions
WARNING: Changes or modifications not expressly approved by Masimo
shall void the warranty for this equipment and could void the user’s
authority to operate the equipment.