Operator's Manual
Root Chapter 14: Service and Maintenance
www.masimo.com 176 Masimo
Cable Type
Nurse Call Event
Menu Setting
3-Circuit
1 and 2 contacts normally
opened
2 and 3 contacts normally
closed
Nurse Call Polarity
Normal
1 and 2 contacts normally
closed
2 and 3 contacts normally
opened
Nurse Call Polarity
Inverse
Battery Test
To conduct a Battery Test
1. Fully charge Root by connecting it to AC power.
2. Verify that the Root Charging Indicator is illuminated.
3. When Root is fully charged, the Root Charging Indicator turns off.
4. Power On Root and verify that the Root Battery Indicator icon on
the Status Bar shows a full charge.
Repair Policy
Masimo or an authorized Service Department must perform warranty repair
and service. Do not use malfunctioning equipment. Have the device
repaired.