Chapter 14: Service and Maintenance
This chapter contains information about cleaning, battery operation,
performance verification, service, repair, and warranty.
Cleaning
Root is a non-sterile and reusable device. The surface of the Root should be
cleaned when the device is visibly dirty, before and after each procedure,
and/or according to hospital practice.
To surface clean, wipe down the outer surface of Root using any of the
following:
Cidex Plus (3.4% glutaraldehyde)
10% bleach solution
70% isopropyl alcohol solution
Do not allow liquids to enter the interior of Root. Using the recommended
cleaning solutions on the touchscreen will not affect the performance of
Root.
Replacing the Fuses
If a power-related problem causes one or both of the fuses to fail, the
fuse(s) will need to be replaced. Replace fuse(s) with UL Listed fuses rated
250V, 2 amp, metric 5x20 mm and with a time delay breaking capacity of
minimum 1500A.
WARNING: To ensure safety, only replace with appropriately rated fuses.
The fuses can be removed by hand or with a 5-millimeter or 3/16-inch
screwdriver.