Operator's Manual
Root Chapter 13: Specifications
www.masimo.com 162 Masimo
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Surge
IEC 61000-4-5
+1 kV -
differential
mode
+2 kV -
common
mode
---
Mains power quality
should be that of a typical
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
100% for 0.5
cycle
60% for 5
cycles
30% for 25
cycles
100% for 5
seconds
---
Mains power quality
should be that of a typical
commercial or hospital
environment. Root
provides a battery for
continued operation
during power mains
interruption for a
maximum of 4 hours.
Power frequency
(50 / 60 Hz)
magnetic field.
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of typical
location in a typical
hospital environment.
Portable and mobile RF communications equipment should be used no
closer to any part of the ME Equipment, including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter.