Operator's Manual
Root Chapter 13: Specifications
www.masimo.com 161 Masimo
Guidance and Manufacturer's
Declaration-Electromagnetic Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the ME Equipment should
assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV
contact
+8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+2 kV for
power supply
lines.
+1 kV for
input/output
lines.
---
Mains power quality
should be that of a typical
commercial or hospital
environment.