Operator's Manual
www.masimo.com 151 Masimo
Chapter 13: Specifications
This chapter contains specifications of Root.
For information on the specifications of Radical-7, Radius-7, MOC-9
modules, and standalone devices, see Directions for Use or Operator's
Manuals for these devices.
Alarms
Audio Alarm
Type
System Status
Light Color
Audio Description
High Priority
Flashing Red
10-pulse burst, pulse spacing: 0.250s,
0.250s, 0.500s, 0.250s, repeat time:10s
Medium
Priority
Flashing Yellow
3-pulse burst, pulse spacing: 0.375s,
0.375s, repeat time: 7s
Low Priority
Solid Yellow
No audio
Nurse Call Specifications
The Nurse Call relays have the following electrical specifications per switch:
Parameter
Specification
Max Voltage
36 VDC or 24 VAC peak