Operator's Manual
Root NIBPT Chapter 3: Operation
www.masimo.com 92 Masimo
Pulse Rate
See
Pulse Rate (PR)
on page 94.
Mean Arterial Pressure
See
Mean Arterial Pressure (MAP)
on page 95.
Systolic/Diastolic (SYS/DIA)
From the Systolic/Diastolic screen, touch Alarms, and then change any of the
following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
Configurable
Options
Systolic
High Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium
220
42-259 in steps
of 1, or Off
When set to Off,
alarm is disabled
Systolic
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium
75
41-258 in steps
of 1, or Off
When set to Off,
alarm is disabled