Operator's Manual
Root NIBPT Chapter 3: Operation
www.masimo.com 85 Masimo
From the Temperature screen, touch Alarms, and then change any of the
following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
Configurable
Options
High Limit
(°F)
The High Limit is
upper threshold that
triggers an alarm.
Medium
Off
80.20-109.9 in
steps of 0.1, or
Off
When set to Off,
alarm is disabled
Low Limit
(°F)
Low Limit is the lower
threshold that
triggers an alarm.
Medium
Off
80.1-109.8 in
steps of 0.1, or
Off
When set to Off,
alarm is disabled
High Limit
(°C)
The High Limit is
upper threshold that
triggers an alarm.
Medium
Off
26.9-43.2 in steps
of 0.1, or Off
When set to Off,
alarm is disabled
Low Limit
(°C)
Low Limit is the lower
threshold that
triggers an alarm.
Medium
Off
26.8-43.1 in steps
of 0.1, or Off
When set to Off,
alarm is disabled